Your ST108 Water Result Is Out of Specification. Now What?

Illustrated ST108 water testing report showing an out-of-specification water quality result in sterile processing

Your facility receives its latest sterile processing water test report, and one or more parameters fall outside the applicable water-quality specifications.

The immediate question is:

What happens when Critical Water fails to meet the required specifications?

An unexpected result can raise concerns about treatment-system performance, medical device processing, and whether corrective action is necessary.

Understanding what the result indicates, how to investigate potential causes, and when follow-up testing is appropriate can help healthcare teams respond with greater confidence.

What Does an Out-of-Specification ST108 Result Mean?

ANSI/AAMI ST108:2023 establishes water-quality requirements for medical device processing, distinguishing between Utility Water, Critical Water, and Steam.

An out-of-specification result occurs when a measured parameter falls outside its applicable acceptance limits. Depending on the water category, these parameters may include conductivity, pH, hardness, chemical contaminants, bacterial counts, and endotoxin levels.

Not every finding indicates the same problem. For example, elevated conductivity may suggest changes in dissolved ionic contaminants, while increased bacterial counts may point toward microbial conditions requiring investigation.

A failed result identifies a condition requiring attention, but it does not automatically explain its cause.

It is also important to distinguish between alert levels and action levels. Under ANSI/AAMI ST108, alert levels provide an early warning that water-quality conditions may be changing, while action levels indicate that immediate intervention is necessary. Understanding which threshold has been reached helps facilities determine the urgency of their response.

Understanding What the Results Reveal

Before determining the next action, review the laboratory report to understand exactly what was measured.

Important considerations include:

  • The parameter: Which chemical or microbiological measurement is outside specification?
  • The result: How does the measured value compare with the applicable limit?
  • The sampling location: Was the sample collected from a treatment outlet, distribution point, or point of use?
  • The reporting details: Were the correct analytical methods, units, sample identifiers, and applicable limits used?

Reviewing previous results can also help determine whether the finding represents an isolated change or part of a developing trend.

What the Finding Means for Sterile Processing

Once the finding is identified, the facility’s responsible personnel should assess its significance for the affected water application.

Sterile processing leadership, infection prevention, facilities engineering, and water treatment specialists may need to participate in this evaluation.

The appropriate response depends on the parameter involved, the severity of the finding, the intended water use, and the facility’s established procedures.

A result affecting water used during final rinsing may require different considerations from one involving water used during initial washing.

The priority is determining whether the affected water remains suitable for its intended use and whether immediate precautions are necessary.

Potentially significant findings should be escalated promptly in accordance with the facility’s procedures.

Where Water Quality Problems Can Originate

Water-quality changes can originate from several points within a facility’s treatment and distribution system.

Potential contributing factors include:

  • Treatment performance: Membrane fouling, exhausted filtration media, or equipment malfunction.
  • Storage and distribution: Biofilm development, stagnant water, or poorly maintained piping.
  • Incoming water quality: Changes in source-water chemistry or treatment demands.
  • Maintenance activities: Recent repairs, system modifications, or changes in operating conditions.

Reviewing maintenance records, treatment-system measurements, and sampling locations alongside laboratory results can help narrow the investigation.

The objective is to identify the underlying issue rather than relying on assumptions about a single failed parameter.

Addressing the Issue and Documenting the Response

Corrective actions should address the conditions identified during the investigation.

Depending on the problem, this may involve equipment maintenance, treatment adjustments, system cleaning or disinfection, or changes to operational procedures.

The appropriate intervention should be determined by the facility’s responsible personnel in accordance with applicable requirements and equipment instructions.

Documentation should capture the original findings, investigation, decisions, corrective measures, and personnel involved.

These records help demonstrate how the facility responded and support future water-quality evaluations.

Follow-Up Testing

Corrective action is not the final step. Follow-up testing helps determine whether the affected water-quality parameters have returned to acceptable levels.

A follow-up sample should support verification of the corrective action, not be used to dismiss the original finding without appropriate investigation.

Testing locations and methods should reflect the original finding and the actions performed.

For example, when elevated bacterial counts are detected at a point of use, testing only at the treatment-system outlet may not adequately verify conditions at the affected location.

Appropriate water sampling locations are therefore important for evaluating corrective-action effectiveness.

The facility should review the follow-up results and determine whether further investigation, monitoring, or operational precautions are necessary.

How I-2-I Solutions Supports ST108 Water Testing

I-2-I Solutions supports healthcare facilities and sterile processing departments through professional water sampling, coordinated laboratory testing, and water-quality reporting.

Our SPS and Critical Water Testing services support routine monitoring, investigations, and follow-up verification of Utility Water, Critical Water, and Steam systems.

Through appropriate sampling and coordinated chemical and microbiological testing, we provide data that helps facilities evaluate water-quality conditions and support informed decisions.

An Out-of-Specification Result Is a Starting Point

Receiving an unexpected ST108 water test result can raise important operational concerns, but understanding what comes next is essential.

Reviewing the finding, evaluating its significance, investigating potential causes, implementing appropriate corrective measures, and verifying water quality are all parts of an effective response.

The goal is not simply to obtain a passing result. It is to understand what changed, address the underlying issue, and maintain confidence in the water used for medical device processing.

References

  1. Association for the Advancement of Medical Instrumentation (AAMI). ANSI/AAMI ST108:2023 - Water for the Processing of Medical Devices.
  2. American National Standards Institute (ANSI). ANSI/AAMI ST108:2023 - Water for Processing Medical Devices.
  3. Association for the Advancement of Medical Instrumentation (AAMI). Corrosion Threatens Sterile Processing-Until Standards Step In.
Healthcare team reviewing a water quality management program